• Promotion
ELISA Kit for Beta-2-Microglobulin (b2M)

BMG; B2MG

Specificity

This assay has high sensitivity and excellent specificity for detection of Beta-2-Microglobulin (b2M).
No significant cross-reactivity or interference between Beta-2-Microglobulin (b2M) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Beta-2-Microglobulin (b2M) and the recovery rates were calculated by comparing the measured value to the expected amount of Beta-2-Microglobulin (b2M) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 93-101 98
EDTA plasma(n=5) 84-96 91
heparin plasma(n=5) 98-105 102

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Beta-2-Microglobulin (b2M) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Beta-2-Microglobulin (b2M) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Beta-2-Microglobulin (b2M) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 84-92% 84-93% 94-105% 84-98%
EDTA plasma(n=5) 83-97% 82-99% 88-96% 98-105%
heparin plasma(n=5) 81-94% 99-105% 92-101% 82-95%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×120µL Assay Diluent A 1×12mL
Detection Reagent B 1×120µL Assay Diluent B 1×12mL
TMB Substrate 1×9mL Stop Solution 1×6mL
Wash Buffer (30 × concentrate) 1×20mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.

GIVEAWAYS

INCREMENT SERVICES

Magazine Reference
BioMed Research International The Renal Effects of Vanadate Exposure: Potential Biomarkers and Oxidative Stress as a Mechanism of Functional Renal Disorders—Preliminary Studies Hindawi: 740105
BMC Nephrology Absence of chloride intracellular channel 4 (CLIC4) predisposes to acute kidney injury but has minimal impact on recovery Biomedcentral: Source
Catalog No. Related products for research use of Homo sapiens (Human) Organism species Applications (RESEARCH USE ONLY!)
RPA260Hu01 Recombinant Beta-2-Microglobulin (b2M) Positive Control; Immunogen; SDS-PAGE; WB.
EPA260Hu61 Eukaryotic Beta-2-Microglobulin (b2M) Positive Control; Immunogen; SDS-PAGE; WB.
PAA260Hu06 Polyclonal Antibody to Beta-2-Microglobulin (b2M) WB; IHC; ICC; IP.
PAA260Hu01 Polyclonal Antibody to Beta-2-Microglobulin (b2M) WB; IHC; ICC; IP.
FAA260Hu01 Anti-Beta-2-Microglobulin (b2M) Polyclonal Antibody Flow cytometry.
LAA260Hu71 Biotin-Linked Polyclonal Antibody to Beta-2-Microglobulin (b2M) WB; IHC; ICC.
LAA260Hu81 FITC-Linked Polyclonal Antibody to Beta-2-Microglobulin (b2M) WB; IHC; ICC; IF.
MAA260Hu22 Monoclonal Antibody to Beta-2-Microglobulin (b2M) WB; IHC; ICC; IP.
SEA260Hu ELISA Kit for Beta-2-Microglobulin (b2M) Enzyme-linked immunosorbent assay for Antigen Detection.
SCA260Hu CLIA Kit for Beta-2-Microglobulin (b2M) Chemiluminescent immunoassay for Antigen Detection.
KSA260Hu01 ELISA Kit DIY Materials for Beta-2-Microglobulin (b2M) Main materials for "Do It (ELISA Kit) Yourself".