ELISA Kit for Ornithine Transcarbamylase (OTC)

OCTD; OCT; OTCase; Ornithine Transcarbamylase; Ornithine Carbamoyltransferase

Specificity

This assay has high sensitivity and excellent specificity for detection of Ornithine Transcarbamylase (OTC).
No significant cross-reactivity or interference between Ornithine Transcarbamylase (OTC) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Ornithine Transcarbamylase (OTC) and the recovery rates were calculated by comparing the measured value to the expected amount of Ornithine Transcarbamylase (OTC) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 94-101 97
EDTA plasma(n=5) 97-104 101
heparin plasma(n=5) 80-98 86

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Ornithine Transcarbamylase (OTC) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Ornithine Transcarbamylase (OTC) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Ornithine Transcarbamylase (OTC) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 96-103% 83-94% 97-105% 96-104%
EDTA plasma(n=5) 78-101% 80-102% 89-96% 84-91%
heparin plasma(n=5) 85-93% 87-103% 83-94% 78-98%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×120µL Assay Diluent A 1×12mL
Detection Reagent B 1×120µL Assay Diluent B 1×12mL
TMB Substrate 1×9mL Stop Solution 1×6mL
Wash Buffer (30 × concentrate) 1×20mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Biochemical and Biophysical Research Communications Circadian oscillation expression of ornithine carbamoyltransferase and its significance in sleep disturbance 33957483
Catalog No. Related products for research use of Rattus norvegicus (Rat) Organism species Applications (RESEARCH USE ONLY!)
RPB492Ra01 Recombinant Ornithine Transcarbamylase (OTC) Positive Control; Immunogen; SDS-PAGE; WB.
PAB492Ra01 Polyclonal Antibody to Ornithine Transcarbamylase (OTC) WB; IHC; ICC; IP.
MAB492Ra21 Monoclonal Antibody to Ornithine Transcarbamylase (OTC) WB; IHC; ICC; IP.
SEB492Ra ELISA Kit for Ornithine Transcarbamylase (OTC) Enzyme-linked immunosorbent assay for Antigen Detection.
LMB492Ra Multiplex Assay Kit for Ornithine Transcarbamylase (OTC) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.