Multiplex Assay Kit for Tissue Polypeptide Antigen (TPA) ,etc. by CBA (Cytometric Bead Array)

Specificity

This assay has high sensitivity and excellent specificity for detection of Tissue Polypeptide Antigen (TPA) ,etc. by CBA (Cytometric Bead Array).
No significant cross-reactivity or interference between Tissue Polypeptide Antigen (TPA) ,etc. by CBA (Cytometric Bead Array) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Tissue Polypeptide Antigen (TPA) ,etc. by CBA (Cytometric Bead Array) and the recovery rates were calculated by comparing the measured value to the expected amount of Tissue Polypeptide Antigen (TPA) ,etc. by CBA (Cytometric Bead Array) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 96-103 101
EDTA plasma(n=5) 94-102 98
heparin plasma(n=5) 79-91 85

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Tissue Polypeptide Antigen (TPA) ,etc. by CBA (Cytometric Bead Array) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Tissue Polypeptide Antigen (TPA) ,etc. by CBA (Cytometric Bead Array) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Tissue Polypeptide Antigen (TPA) ,etc. by CBA (Cytometric Bead Array) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 98-105% 95-104% 86-94% 80-88%
EDTA plasma(n=5) 81-89% 99-105% 78-103% 92-104%
heparin plasma(n=5) 89-99% 91-104% 80-94% 94-101%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity

Assay procedure summary

1. Add 200μL analysis buffer solution to each well of the plate for pre-wetting;
2. Preparation of standards, reagents and samples before the experiment;
3. Add 100μL standard or sample to each well, add 10μL magnetic beads, and incubate 120 minutes protect from light at 25°C on shaker;
4. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60 minutes protect from light at 25°C on shaker;
5. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent B, and incubate 30 minutes protect from light at 25°C on shaker;
6. Wash plate once on magnetic frame;
7. Add 100μL sheath fluid, swirl for 10 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Lipids in Health and disease Biofunctional nanogold microsphere doped with Prussian blue nanoparticles for sensitive electrochemical immunoassay of cancer marker NCBI: PMC2914008
Journal of Cancer Therapy Diagnostic and Prognostic Value of Serum Tissue Polypeptide Antigen in Advanced Lung Cancer Scirp: Source
39 Cytokine and Estrogen Stimulation of Endothelial Cells Augments Activation of the Surface-Bound Prekallikrein-High Molecular Weight Kininogen Complex: Implications for Hereditary Angioedema (HAE): 814 article:S0091-6749(15)02643-3
Annals of Clinical & Laboratory Science Alteration of Hemostatic Parameters in Patients with Different Levels of Subclinical Hypothyroidism and the Effect of L-thyroxine Treatment pubmed:28249913
Contraception Antifertility effectiveness of a novel polymer matrix composite and its influence on the endometrium in rhesus macaques (Macaca mulatta) Pubmed: 30980831
Catalog No. Related products for research use of Homo sapiens (Human) Organism species Applications (RESEARCH USE ONLY!)
SEA163Hu ELISA Kit for Tissue Polypeptide Antigen (TPA) Enzyme-linked immunosorbent assay for Antigen Detection.
CMA163Hu Multiplex Assay Kit for Tissue Polypeptide Antigen (TPA) ,etc. by CBA (Cytometric Bead Array) FLIA Kit for Antigen Detection.
LMA163Hu Multiplex Assay Kit for Tissue Polypeptide Antigen (TPA) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.