Multiplex Assay Kit for Desmin (Des) ,etc. by CBA (Cytometric Bead Array)

CMD1I; CSM1; CSM2; Intermediate Filament Protein

Specificity

This assay has high sensitivity and excellent specificity for detection of Desmin (Des) ,etc. by CBA (Cytometric Bead Array).
No significant cross-reactivity or interference between Desmin (Des) ,etc. by CBA (Cytometric Bead Array) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Desmin (Des) ,etc. by CBA (Cytometric Bead Array) and the recovery rates were calculated by comparing the measured value to the expected amount of Desmin (Des) ,etc. by CBA (Cytometric Bead Array) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 85-94 88
EDTA plasma(n=5) 78-101 84
heparin plasma(n=5) 97-104 101

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Desmin (Des) ,etc. by CBA (Cytometric Bead Array) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Desmin (Des) ,etc. by CBA (Cytometric Bead Array) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Desmin (Des) ,etc. by CBA (Cytometric Bead Array) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 95-102% 97-105% 78-99% 78-94%
EDTA plasma(n=5) 80-103% 78-90% 95-102% 92-99%
heparin plasma(n=5) 89-105% 89-96% 84-97% 78-94%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity

Assay procedure summary

1. Add 200μL analysis buffer solution to each well of the plate for pre-wetting;
2. Preparation of standards, reagents and samples before the experiment;
3. Add 100μL standard or sample to each well, add 10μL magnetic beads, and incubate 120 minutes protect from light at 25°C on shaker;
4. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60 minutes protect from light at 25°C on shaker;
5. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent B, and incubate 30 minutes protect from light at 25°C on shaker;
6. Wash plate once on magnetic frame;
7. Add 100μL sheath fluid, swirl for 10 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
PLoS One. Utility of Survival Motor Neuron ELISA for Spinal Muscular Atrophy Clinical and Preclinical Analyses PubMed: PMC3164180
Dis Markers. A Novel Panel of Serum Biomarkers for MPM Diagnosis. pubmed:28348450
Catalog No. Related products for research use of Rattus norvegicus (Rat) Organism species Applications (RESEARCH USE ONLY!)
RPA373Ra01 Recombinant Desmin (Des) Positive Control; Immunogen; SDS-PAGE; WB.
PAA373Ra01 Polyclonal Antibody to Desmin (Des) Testing in progress
MAA373Ra21 Monoclonal Antibody to Desmin (Des) WB; IHC; ICC; IP.
SEA373Ra ELISA Kit for Desmin (Des) Enzyme-linked immunosorbent assay for Antigen Detection.
CMA373Ra Multiplex Assay Kit for Desmin (Des) ,etc. by CBA (Cytometric Bead Array) FLIA Kit for Antigen Detection.
LMA373Ra Multiplex Assay Kit for Desmin (Des) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.