Multiplex Assay Kit for Estriol (E3) ,etc. by CBA (Cytometric Bead Array)

oestriol

Specificity

This assay has high sensitivity and excellent specificity for detection of Estriol (E3) ,etc. by CBA (Cytometric Bead Array).
No significant cross-reactivity or interference between Estriol (E3) ,etc. by CBA (Cytometric Bead Array) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of Estriol (E3) ,etc. by CBA (Cytometric Bead Array) and the recovery rates were calculated by comparing the measured value to the expected amount of Estriol (E3) ,etc. by CBA (Cytometric Bead Array) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 95-104 101
EDTA plasma(n=5) 83-90 86
heparin plasma(n=5) 88-104 99

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Estriol (E3) ,etc. by CBA (Cytometric Bead Array) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Estriol (E3) ,etc. by CBA (Cytometric Bead Array) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Estriol (E3) ,etc. by CBA (Cytometric Bead Array) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 82-98% 80-101% 89-98% 90-97%
EDTA plasma(n=5) 88-95% 97-104% 81-92% 95-102%
heparin plasma(n=5) 87-101% 84-105% 97-105% 96-103%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
96-well plate 1 Plate sealer for 96 wells 4
Pre-Mixed Standard 2 Standard Diluent 1×20mL
Pre-Mixed Magnetic beads (22#:E3) 1 Analysis buffer 1×20mL
Pre-Mixed Detection Reagent A 1×120μL Assay Diluent A 1×12mL
Detection Reagent B (PE-SA) 1×120μL Assay Diluent B 1×12mL
Sheath Fluid 1×10mL Wash Buffer (30 × concentrate) 1×20mL
Instruction manual 1

Assay procedure summary

1. Add 200μL analysis buffer solution to each well of the plate for pre-wetting;
2. Preparation of standards, reagents and samples before the experiment;
3. Add 50μL standard or sample to each well, add 50μL prepared Detection Reagent A, add 10μL magnetic beads, and incubate 90 minutes protect from light at 25°C on shaker;
4. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent B, and incubate 30 minutes protect from light at 25°C on shaker;
5. Wash plate once on magnetic frame;
6. Add 100μL sheath fluid, swirl for 10 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
European journal of pharmacology Osteoprotective effects of salidroside in ovariectomized mice and diabetic mice pubmed:29242120
Catalog No. Related products for research use of Pan-species (General) Organism species Applications (RESEARCH USE ONLY!)
CPA455Ge11 BSA Conjugated Estriol (E3) Immunogen; SDS-PAGE; WB.
CPA455Ge21 OVA Conjugated Estriol (E3) Immunogen; SDS-PAGE; WB.
CPA455Ge31 KLH Conjugated Estriol (E3) Immunogen; SDS-PAGE; WB.
PAA455Ge01 Polyclonal Antibody to Estriol (E3) ELISA
MAA455Ge22 Monoclonal Antibody to Estriol (E3) ELISA, CLIA. / IHC-Fr, ICC, IP (predicted).
MAA455Ge21 Monoclonal Antibody to Estriol (E3) ELISA
CEA455Ge ELISA Kit for Estriol (E3) Enzyme-linked immunosorbent assay for Antigen Detection.
IEA455Ge Instant ELISA Kit for Estriol (E3) Enzyme-linked immunosorbent assay for Antigen Detection.
CCA455Ge CLIA Kit for Estriol (E3) Chemiluminescent immunoassay for Antigen Detection.
LMA455Ge Multiplex Assay Kit for Estriol (E3) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
CMA455Ge Multiplex Assay Kit for Estriol (E3) ,etc. by CBA (Cytometric Bead Array) FLIA Kit for Antigen Detection.
KSA455Ge11 ELISA Kit DIY Materials for Estriol (E3) Main materials for "Do It (ELISA Kit) Yourself".