Multiplex Assay Kit for Lipoprotein, a (Lpa) ,etc. by CBA (Cytometric Bead Array)

Lp(a); AK38; APOA; LP; Sinking Pre-β-Lipoprotein

Specificity

This assay has high sensitivity and excellent specificity for detection of Lipoprotein, a (Lpa) ,etc. by CBA (Cytometric Bead Array).
No significant cross-reactivity or interference between Lipoprotein, a (Lpa) ,etc. by CBA (Cytometric Bead Array) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Lipoprotein, a (Lpa) ,etc. by CBA (Cytometric Bead Array) and the recovery rates were calculated by comparing the measured value to the expected amount of Lipoprotein, a (Lpa) ,etc. by CBA (Cytometric Bead Array) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 84-99 89
EDTA plasma(n=5) 89-102 95
heparin plasma(n=5) 88-98 91

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Lipoprotein, a (Lpa) ,etc. by CBA (Cytometric Bead Array) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Lipoprotein, a (Lpa) ,etc. by CBA (Cytometric Bead Array) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Lipoprotein, a (Lpa) ,etc. by CBA (Cytometric Bead Array) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 84-98% 87-95% 81-93% 78-105%
EDTA plasma(n=5) 81-94% 82-90% 80-103% 96-103%
heparin plasma(n=5) 91-105% 80-95% 93-101% 97-105%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity

Assay procedure summary

1. Add 200μL analysis buffer solution to each well of the plate for pre-wetting;
2. Preparation of standards, reagents and samples before the experiment;
3. Add 100μL standard or sample to each well, add 10μL magnetic beads, and incubate 120 minutes protect from light at 25°C on shaker;
4. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60 minutes protect from light at 25°C on shaker;
5. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent B, and incubate 30 minutes protect from light at 25°C on shaker;
6. Wash plate once on magnetic frame;
7. Add 100μL sheath fluid, swirl for 10 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Molecular Sciences Pigment Epithelium-Derived Factor (PEDF) Protects Osteoblastic Cell Line from Glucocorticoid-Induced Apoptosis via PEDF-R Pubmed:27187377
Evidence-Based Complementary and Alternative Medicine Danshen (Salvia miltiorrhiza) Compounds Improve the Biochemical Indices of the Patients with Coronary HeartDisease. Pubmed:27366196
INTERNATIONAL ARCHIVES OF ALLERGY AND IMMUNOLOGY Lysophosphatidic Acid Regulates the Differentiation of Th2 Cells and Its Antagonist Suppresses Allergic Airway Inflammation Pubmed: 32846422
BMC Gastroenterol Evaluation of serum ATX and LPA as potential diagnostic biomarkers in patients with pancreatic cancer 33568105
blood advances The binding of autotaxin to integrins mediates hyperhomocysteinemia-potentiated platelet activation and thrombosis 34559203
Catalog No. Related products for research use of Homo sapiens (Human) Organism species Applications (RESEARCH USE ONLY!)
APA842Hu01 Active Lipoprotein, a (Lpa) Cell culture; Activity Assays.
RPA842Hu01 Recombinant Lipoprotein, a (Lpa) Positive Control; Immunogen; SDS-PAGE; WB.
PAA842Hu01 Polyclonal Antibody to Lipoprotein, a (Lpa) WB; IHC
MAA842Hu22 Monoclonal Antibody to Lipoprotein, a (Lpa) WB; IHC; ICC; IP.
SEA842Hu ELISA Kit for Lipoprotein, a (Lpa) Enzyme-linked immunosorbent assay for Antigen Detection.
LMA842Hu Multiplex Assay Kit for Lipoprotein, a (Lpa) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
CMA842Hu Multiplex Assay Kit for Lipoprotein, a (Lpa) ,etc. by CBA (Cytometric Bead Array) FLIA Kit for Antigen Detection.