Multiplex Assay Kit for L1-Cell Adhesion Molecule (L1CAM) ,etc. by CBA (Cytometric Bead Array)

CD171; L1-CAM; CAML1; HSAS; HSAS1; MASA; MIC5; N-CAML1; S10; SPG1; Antigen Identified By Monoclonal Antibody R1

Specificity

This assay has high sensitivity and excellent specificity for detection of L1-Cell Adhesion Molecule (L1CAM) ,etc. by CBA (Cytometric Bead Array).
No significant cross-reactivity or interference between L1-Cell Adhesion Molecule (L1CAM) ,etc. by CBA (Cytometric Bead Array) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant L1-Cell Adhesion Molecule (L1CAM) ,etc. by CBA (Cytometric Bead Array) and the recovery rates were calculated by comparing the measured value to the expected amount of L1-Cell Adhesion Molecule (L1CAM) ,etc. by CBA (Cytometric Bead Array) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 84-101 91
EDTA plasma(n=5) 87-103 99
heparin plasma(n=5) 92-99 96

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level L1-Cell Adhesion Molecule (L1CAM) ,etc. by CBA (Cytometric Bead Array) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level L1-Cell Adhesion Molecule (L1CAM) ,etc. by CBA (Cytometric Bead Array) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of L1-Cell Adhesion Molecule (L1CAM) ,etc. by CBA (Cytometric Bead Array) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 87-95% 90-97% 95-104% 82-95%
EDTA plasma(n=5) 81-102% 94-102% 87-103% 84-91%
heparin plasma(n=5) 90-102% 78-91% 92-105% 79-89%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity

Assay procedure summary

1. Add 200μL analysis buffer solution to each well of the plate for pre-wetting;
2. Preparation of standards, reagents and samples before the experiment;
3. Add 100μL standard or sample to each well, add 10μL magnetic beads, and incubate 120 minutes protect from light at 25°C on shaker;
4. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60 minutes protect from light at 25°C on shaker;
5. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent B, and incubate 30 minutes protect from light at 25°C on shaker;
6. Wash plate once on magnetic frame;
7. Add 100μL sheath fluid, swirl for 10 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Arch Gynecol Obstet. The sL1CAM in sera of patients with endometrial and ovarian cancers. pubmed:27832351
Journal of Neurotrauma L1CAM overexpression downregulates phosphacan and upregulates structural plasticity-related genes rostral and caudal to the complete spinal cord transection Pubmed: 31426714
Cancers Extracellular Vesicles-Based Biomarkers Represent a Promising Liquid Biopsy in Endometrial Cancer Pubmed: 31842290
Catalog No. Related products for research use of Mus musculus (Mouse) Organism species Applications (RESEARCH USE ONLY!)
RPA959Mu01 Recombinant L1-Cell Adhesion Molecule (L1CAM) Positive Control; Immunogen; SDS-PAGE; WB.
RPA959Mu02 Recombinant L1-Cell Adhesion Molecule (L1CAM) Positive Control; Immunogen; SDS-PAGE; WB.
PAA959Mu02 Polyclonal Antibody to L1-Cell Adhesion Molecule (L1CAM) Testing in progress.
PAA959Mu01 Polyclonal Antibody to L1-Cell Adhesion Molecule (L1CAM) WB
SEA959Mu ELISA Kit for L1-Cell Adhesion Molecule (L1CAM) Enzyme-linked immunosorbent assay for Antigen Detection.
CMA959Mu Multiplex Assay Kit for L1-Cell Adhesion Molecule (L1CAM) ,etc. by CBA (Cytometric Bead Array) FLIA Kit for Antigen Detection.
LMA959Mu Multiplex Assay Kit for L1-Cell Adhesion Molecule (L1CAM) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.