Multiplex Assay Kit for Cadherin 17 (CDH17) ,etc. by CBA (Cytometric Bead Array)

CDH16; HPT-1; HPT1; LI-Cadherin; Cadherin,Liver Intestine; Intestinal peptide-associated transporter HPT-1; Liver-intestine cadherin

Specificity

This assay has high sensitivity and excellent specificity for detection of Cadherin 17 (CDH17) ,etc. by CBA (Cytometric Bead Array).
No significant cross-reactivity or interference between Cadherin 17 (CDH17) ,etc. by CBA (Cytometric Bead Array) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Cadherin 17 (CDH17) ,etc. by CBA (Cytometric Bead Array) and the recovery rates were calculated by comparing the measured value to the expected amount of Cadherin 17 (CDH17) ,etc. by CBA (Cytometric Bead Array) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 88-101 97
EDTA plasma(n=5) 78-98 91
heparin plasma(n=5) 79-103 85

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Cadherin 17 (CDH17) ,etc. by CBA (Cytometric Bead Array) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Cadherin 17 (CDH17) ,etc. by CBA (Cytometric Bead Array) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Cadherin 17 (CDH17) ,etc. by CBA (Cytometric Bead Array) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 91-99% 97-104% 81-89% 84-101%
EDTA plasma(n=5) 88-104% 78-92% 92-99% 93-101%
heparin plasma(n=5) 82-98% 94-103% 81-93% 91-101%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity

Assay procedure summary

1. Add 200μL analysis buffer solution to each well of the plate for pre-wetting;
2. Preparation of standards, reagents and samples before the experiment;
3. Add 100μL standard or sample to each well, add 10μL magnetic beads, and incubate 120 minutes protect from light at 25°C on shaker;
4. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60 minutes protect from light at 25°C on shaker;
5. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent B, and incubate 30 minutes protect from light at 25°C on shaker;
6. Wash plate once on magnetic frame;
7. Add 100μL sheath fluid, swirl for 10 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Urologic oncology sE-cadherin is upregulated in serum of patients with renal cell carcinoma and promotes tumor cell dissemination in vitro Pubmed: 31005422
Catalog No. Related products for research use of Homo sapiens (Human) Organism species Applications (RESEARCH USE ONLY!)
RPC002Hu01 Recombinant Cadherin 17 (CDH17) Positive Control; Immunogen; SDS-PAGE; WB.
PAC002Hu01 Polyclonal Antibody to Cadherin 17 (CDH17) WB; IHC
MAC002Hu22 Monoclonal Antibody to Cadherin 17 (CDH17) WB; IHC; ICC; IP.
SEC002Hu ELISA Kit for Cadherin 17 (CDH17) Enzyme-linked immunosorbent assay for Antigen Detection.
LMC002Hu Multiplex Assay Kit for Cadherin 17 (CDH17) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
CMC002Hu Multiplex Assay Kit for Cadherin 17 (CDH17) ,etc. by CBA (Cytometric Bead Array) FLIA Kit for Antigen Detection.