Multiplex Assay Kit for Reelin (RELN) ,etc. by CBA (Cytometric Bead Array)

RL

Specificity

This assay has high sensitivity and excellent specificity for detection of Reelin (RELN) ,etc. by CBA (Cytometric Bead Array).
No significant cross-reactivity or interference between Reelin (RELN) ,etc. by CBA (Cytometric Bead Array) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Reelin (RELN) ,etc. by CBA (Cytometric Bead Array) and the recovery rates were calculated by comparing the measured value to the expected amount of Reelin (RELN) ,etc. by CBA (Cytometric Bead Array) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 98-105 102
EDTA plasma(n=5) 96-103 99
heparin plasma(n=5) 91-101 98

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Reelin (RELN) ,etc. by CBA (Cytometric Bead Array) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Reelin (RELN) ,etc. by CBA (Cytometric Bead Array) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Reelin (RELN) ,etc. by CBA (Cytometric Bead Array) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 79-94% 79-99% 79-94% 79-96%
EDTA plasma(n=5) 88-101% 86-93% 87-98% 78-92%
heparin plasma(n=5) 99-105% 79-101% 81-102% 82-89%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity

Assay procedure summary

1. Add 200μL analysis buffer solution to each well of the plate for pre-wetting;
2. Preparation of standards, reagents and samples before the experiment;
3. Add 100μL standard or sample to each well, add 10μL magnetic beads, and incubate 120 minutes protect from light at 25°C on shaker;
4. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60 minutes protect from light at 25°C on shaker;
5. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent B, and incubate 30 minutes protect from light at 25°C on shaker;
6. Wash plate once on magnetic frame;
7. Add 100μL sheath fluid, swirl for 10 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Scientific Reports The β-amyloid peptide compromises Reelin signaling in Alzheimer's disease pubmed:27531658
clinical chemistry and laboratory medicine Blood reelin levels in the progression of chronic liver disease Pubmed: 24803609
Adv Med Sci Blood reelin in the progression of chronic liver disease 33561810
Biomolecules Surfactant Protein-G in Wildtype and 3xTg-AD Mice: Localization in the Forebrain, Age-Dependent Hippocampal Dot-like Deposits and Brain Content Pubmed:35053244
Catalog No. Related products for research use of Mus musculus (Mouse) Organism species Applications (RESEARCH USE ONLY!)
RPC775Mu01 Recombinant Reelin (RELN) Positive Control; Immunogen; SDS-PAGE; WB.
RPC775Mu02 Recombinant Reelin (RELN) Positive Control; Immunogen; SDS-PAGE; WB.
PAC775Mu01 Polyclonal Antibody to Reelin (RELN) WB,IHC,ICC/IF
PAC775Mu02 Polyclonal Antibody to Reelin (RELN) WB; IHC; ICC; IP.
MEC775Mu Mini Samples ELISA Kit for Reelin (RELN) Enzyme-linked immunosorbent assay for Antigen Detection.
SEC775Mu ELISA Kit for Reelin (RELN) Enzyme-linked immunosorbent assay for Antigen Detection.
CMC775Mu Multiplex Assay Kit for Reelin (RELN) ,etc. by CBA (Cytometric Bead Array) FLIA Kit for Antigen Detection.
LMC775Mu Multiplex Assay Kit for Reelin (RELN) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.