Multiplex Assay Kit for Syntenin 2 (ST2) ,etc. by CBA (Cytometric Bead Array)

SDCBP2; SITAC18; Syndecan Binding Protein 2

Specificity

This assay has high sensitivity and excellent specificity for detection of Syntenin 2 (ST2) ,etc. by CBA (Cytometric Bead Array).
No significant cross-reactivity or interference between Syntenin 2 (ST2) ,etc. by CBA (Cytometric Bead Array) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Syntenin 2 (ST2) ,etc. by CBA (Cytometric Bead Array) and the recovery rates were calculated by comparing the measured value to the expected amount of Syntenin 2 (ST2) ,etc. by CBA (Cytometric Bead Array) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 85-96 89
EDTA plasma(n=5) 90-99 95
heparin plasma(n=5) 80-101 90

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Syntenin 2 (ST2) ,etc. by CBA (Cytometric Bead Array) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Syntenin 2 (ST2) ,etc. by CBA (Cytometric Bead Array) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Syntenin 2 (ST2) ,etc. by CBA (Cytometric Bead Array) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 91-101% 84-105% 83-97% 79-98%
EDTA plasma(n=5) 91-105% 95-104% 97-104% 96-104%
heparin plasma(n=5) 84-93% 89-101% 99-105% 86-95%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
96-well plate 1 Plate sealer for 96 wells 4
Pre-Mixed Standard 2 Standard Diluent 1×20mL
Pre-Mixed Magnetic beads (22#:ST2) 1 Analysis buffer 1×20mL
Pre-Mixed Detection Reagent A 1×120μL Assay Diluent A 1×12mL
Detection Reagent B (PE-SA) 1×120μL Assay Diluent B 1×12mL
Sheath Fluid 1×10mL Wash Buffer (30 × concentrate) 1×20mL
Instruction manual 1

Assay procedure summary

1. Add 200μL analysis buffer solution to each well of the plate for pre-wetting;
2. Preparation of standards, reagents and samples before the experiment;
3. Add 100μL standard or sample to each well, add 10μL magnetic beads, and incubate 120 minutes protect from light at 25°C on shaker;
4. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60 minutes protect from light at 25°C on shaker;
5. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent B, and incubate 30 minutes protect from light at 25°C on shaker;
6. Wash plate once on magnetic frame;
7. Add 100μL sheath fluid, swirl for 10 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Bioscience Reports Inhibition of microRNA-146a attenuated heart failure in myocardial infarction rats Pubmed: 31763669
Microbiome changes in patients with chronic heart failure with preserved ejection fraction correlate with fibrosis markers: Description of a Russian cohort
J Immunol Res Inhibition of NF-κB/IL-33/ST2 Axis Ameliorates Acute Bronchiolitis Induced by Respiratory Syncytial Virus 34395633
Catalog No. Related products for research use of Homo sapiens (Human) Organism species Applications (RESEARCH USE ONLY!)
RPD171Hu01 Recombinant Syntenin 2 (ST2) Positive Control; Immunogen; SDS-PAGE; WB.
PAD171Hu01 Polyclonal Antibody to Syntenin 2 (ST2) WB,IHC,ICC/IF
MAD171Hu21 Monoclonal Antibody to Syntenin 2 (ST2) WB,IHC
SED171Hu ELISA Kit for Syntenin 2 (ST2) Enzyme-linked immunosorbent assay for Antigen Detection.
CMD171Hu Multiplex Assay Kit for Syntenin 2 (ST2) ,etc. by CBA (Cytometric Bead Array) FLIA Kit for Antigen Detection.
LMD171Hu Multiplex Assay Kit for Syntenin 2 (ST2) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
PGD171Hu01 Primer Pair for Syntenin 2 (ST2) PCR