Multiplex Assay Kit for Serine/Threonine Kinase 39 (STK39) ,etc. by CBA (Cytometric Bead Array)

DCHT; SPAK; STE20/SPS1-related proline-alanine-rich protein kinase

Specificity

This assay has high sensitivity and excellent specificity for detection of Serine/Threonine Kinase 39 (STK39) ,etc. by CBA (Cytometric Bead Array).
No significant cross-reactivity or interference between Serine/Threonine Kinase 39 (STK39) ,etc. by CBA (Cytometric Bead Array) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Serine/Threonine Kinase 39 (STK39) ,etc. by CBA (Cytometric Bead Array) and the recovery rates were calculated by comparing the measured value to the expected amount of Serine/Threonine Kinase 39 (STK39) ,etc. by CBA (Cytometric Bead Array) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 92-99 95
EDTA plasma(n=5) 78-97 84
heparin plasma(n=5) 95-103 101

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Serine/Threonine Kinase 39 (STK39) ,etc. by CBA (Cytometric Bead Array) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Serine/Threonine Kinase 39 (STK39) ,etc. by CBA (Cytometric Bead Array) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Serine/Threonine Kinase 39 (STK39) ,etc. by CBA (Cytometric Bead Array) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 78-93% 84-104% 85-95% 81-92%
EDTA plasma(n=5) 78-96% 93-101% 88-95% 92-101%
heparin plasma(n=5) 78-105% 80-92% 92-101% 78-97%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity

Assay procedure summary

1. Add 200μL analysis buffer solution to each well of the plate for pre-wetting;
2. Preparation of standards, reagents and samples before the experiment;
3. Add 100μL standard or sample to each well, add 10μL magnetic beads, and incubate 120 minutes protect from light at 25°C on shaker;
4. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60 minutes protect from light at 25°C on shaker;
5. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent B, and incubate 30 minutes protect from light at 25°C on shaker;
6. Wash plate once on magnetic frame;
7. Add 100μL sheath fluid, swirl for 10 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Catalog No. Related products for research use of Homo sapiens (Human) Organism species Applications (RESEARCH USE ONLY!)
RPN826Hu01 Recombinant Serine/Threonine Kinase 39 (STK39) Positive Control; Immunogen; SDS-PAGE; WB.
PAN826Hu01 Polyclonal Antibody to Serine/Threonine Kinase 39 (STK39) IHC
MAN826Hu21 Monoclonal Antibody to Serine/Threonine Kinase 39 (STK39) WB; IHC; ICC; IP.
SEN826Hu ELISA Kit for Serine/Threonine Kinase 39 (STK39) Enzyme-linked immunosorbent assay for Antigen Detection.
SCN826Hu CLIA Kit for Serine/Threonine Kinase 39 (STK39) Chemiluminescent immunoassay for Antigen Detection.
CMN826Hu Multiplex Assay Kit for Serine/Threonine Kinase 39 (STK39) ,etc. by CBA (Cytometric Bead Array) FLIA Kit for Antigen Detection.
LMN826Hu Multiplex Assay Kit for Serine/Threonine Kinase 39 (STK39) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.