Multiplex Assay Kit for Ficolin 3 (FCN3) ,etc. by FLIA (Flow Luminescence Immunoassay)

FCNH; HAKA1; Collagen/Fibrinogen Domain Containing; Hakata Antigen; Collagen/fibrinogen domain-containing lectin 3 p35

(Note: Up to 8-plex in one testing reaction)

Specificity

This assay has high sensitivity and excellent specificity for detection of Ficolin 3 (FCN3) ,etc. by FLIA (Flow Luminescence Immunoassay).
No significant cross-reactivity or interference between Ficolin 3 (FCN3) ,etc. by FLIA (Flow Luminescence Immunoassay) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Ficolin 3 (FCN3) ,etc. by FLIA (Flow Luminescence Immunoassay) and the recovery rates were calculated by comparing the measured value to the expected amount of Ficolin 3 (FCN3) ,etc. by FLIA (Flow Luminescence Immunoassay) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 95-102 99
EDTA plasma(n=5) 83-98 94
heparin plasma(n=5) 86-97 90

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Ficolin 3 (FCN3) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Ficolin 3 (FCN3) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Ficolin 3 (FCN3) ,etc. by FLIA (Flow Luminescence Immunoassay) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 93-101% 83-99% 79-104% 94-102%
EDTA plasma(n=5) 79-89% 83-94% 96-105% 81-105%
heparin plasma(n=5) 80-101% 87-103% 97-105% 82-91%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
96-well plate 1 Plate sealer for 96 wells 4
Pre-Mixed Standard 2 Standard Diluent 1×20mL
Pre-Mixed Magnetic beads (22#:FCN3) 1 Analysis buffer 1×20mL
Pre-Mixed Detection Reagent A 1×120μL Assay Diluent A 1×12mL
Detection Reagent B (PE-SA) 1×120μL Assay Diluent B 1×12mL
Sheath Fluid 1×10mL Wash Buffer (30 × concentrate) 1×20mL
Instruction manual 1

Assay procedure summary

1. Preparation of standards, reagents and samples before the experiment;
2. Add 100μL standard or sample to each well,
    add 10μL magnetic beads, and incubate 90min at 37°C on shaker;
3. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60min at 37°C on shaker;
4. Wash plate on magnetic frame for three times;
5. Add 100μL prepared Detection Reagent B, and incubate 30 min at 37°C on shaker;
6. Wash plate on magnetic frame for three times;
7. Add 100μL sheath solution, swirl for 2 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
American Journal of Ophthalmology Correlation between Ficolin-3 and Vascular Endothelial Growth Factor-to-Pigment Epithelium-Derived Factor Ratio in the Vitreous of Eyes with Proliferative Diabetic Retinopathy ScienceDirect: S0002939411004491
Liver International Identification of candidate biomarkers for hepatocellular carcinoma in plasma of HCV-infected cirrhotic patients by 5-D DIGE Pubmed: 23944848
Fundamental & Clinical Pharmacology The determination of histopathological and biochemical effects of the rabbit knee joint injected dexketoprofen trometamol Onlinelibrary: fcp.12074
The Annals of Thoracic Surgery Proteomic Study Reveals Plasma Protein Changes in Congenital Heart Diseases Pubmed: 24565402
Journal of Proteome Research Targeted proteomics predict a sustained complete-response after transarterial chemoembolization and clinical outcomes in patients with hepatocellular carcinoma: a prospective cohort study pubmed:28112944
BMC Cancer Screening for immune-potentiating antigens from hepatocellular carcinoma patients after radiofrequency ablation by serum proteomic analysis Pubmed:29386009
ACTA TROPICA Novel findings on the role of ficolins and colectins in the innate response against Leishmania braziliensis Pubmed: 32827456
Diagnostics Systemic Alterations of Immune Response-Related Proteins during Glaucoma Development in the Murine Model DBA/2J Pubmed: 32585848
Immunol Lett Ficolin-3 in Rheumatic Fever and Rheumatic Heart Disease 33232720