Multiplex Assay Kit for Interleukin 11 (IL11) ,etc. by CBA (Cytometric Bead Array)

AGIF; Adipogenesis Inhibitory Factor; Oprelvekin

Specificity

This assay has high sensitivity and excellent specificity for detection of Interleukin 11 (IL11) ,etc. by CBA (Cytometric Bead Array).
No significant cross-reactivity or interference between Interleukin 11 (IL11) ,etc. by CBA (Cytometric Bead Array) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Interleukin 11 (IL11) ,etc. by CBA (Cytometric Bead Array) and the recovery rates were calculated by comparing the measured value to the expected amount of Interleukin 11 (IL11) ,etc. by CBA (Cytometric Bead Array) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 91-105 95
EDTA plasma(n=5) 80-98 93
heparin plasma(n=5) 80-98 88

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Interleukin 11 (IL11) ,etc. by CBA (Cytometric Bead Array) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Interleukin 11 (IL11) ,etc. by CBA (Cytometric Bead Array) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Interleukin 11 (IL11) ,etc. by CBA (Cytometric Bead Array) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 94-105% 81-97% 89-101% 81-105%
EDTA plasma(n=5) 86-94% 83-103% 94-102% 83-91%
heparin plasma(n=5) 80-93% 79-89% 97-105% 85-99%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity

Assay procedure summary

1. Add 200μL analysis buffer solution to each well of the plate for pre-wetting;
2. Preparation of standards, reagents and samples before the experiment;
3. Add 100μL standard or sample to each well, add 10μL magnetic beads, and incubate 120 minutes protect from light at 25°C on shaker;
4. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60 minutes protect from light at 25°C on shaker;
5. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent B, and incubate 30 minutes protect from light at 25°C on shaker;
6. Wash plate once on magnetic frame;
7. Add 100μL sheath fluid, swirl for 10 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Molecular Medicine Reports Regulatory effect of chemerin and therapeutic efficacy of chemerin‑9 in pancreatogenic diabetes mellitus Pubmed: 31922216
Catalog No. Related products for research use of Homo sapiens (Human) Organism species Applications (RESEARCH USE ONLY!)
EPA057Hu61 Eukaryotic Interleukin 11 (IL11) Positive Control; Immunogen; SDS-PAGE; WB.
APA057Hu01 Active Interleukin 11 (IL11) Cell culture; Activity Assays.
EPA057Hu62 Eukaryotic Interleukin 11 (IL11) Positive Control; Immunogen; SDS-PAGE; WB.
RPA057Hu01 Recombinant Interleukin 11 (IL11) Positive Control; Immunogen; SDS-PAGE; WB.
APA057Hu62 Active Interleukin 11 (IL11) Cell culture; Activity Assays.
APA057Hu61 Active Interleukin 11 (IL11) Cell culture; Activity Assays.
PAA057Hu01 Polyclonal Antibody to Interleukin 11 (IL11) WB
MAA057Hu22 Monoclonal Antibody to Interleukin 11 (IL11) WB; IHC; ICC; IP.
MAA057Hu21 Monoclonal Antibody to Interleukin 11 (IL11) WB
SEA057Hu ELISA Kit for Interleukin 11 (IL11) Enzyme-linked immunosorbent assay for Antigen Detection.
CMA057Hu Multiplex Assay Kit for Interleukin 11 (IL11) ,etc. by CBA (Cytometric Bead Array) FLIA Kit for Antigen Detection.
LMA057Hu Multiplex Assay Kit for Interleukin 11 (IL11) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.