Multiplex Assay Kit for Folic Acid (FA) ,etc. by CBA (Cytometric Bead Array)

VB9; Vitamin B9; Folacin; Folate; V; Itamin M; Vitamin M; Vitamin Bc; Pteroyl-L-Glutamic Acid; Pteroyl-L-Glutamate

Specificity

This assay has high sensitivity and excellent specificity for detection of Folic Acid (FA) ,etc. by CBA (Cytometric Bead Array).
No significant cross-reactivity or interference between Folic Acid (FA) ,etc. by CBA (Cytometric Bead Array) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of Folic Acid (FA) ,etc. by CBA (Cytometric Bead Array) and the recovery rates were calculated by comparing the measured value to the expected amount of Folic Acid (FA) ,etc. by CBA (Cytometric Bead Array) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 83-97 92
EDTA plasma(n=5) 98-105 102
heparin plasma(n=5) 84-98 90

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Folic Acid (FA) ,etc. by CBA (Cytometric Bead Array) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Folic Acid (FA) ,etc. by CBA (Cytometric Bead Array) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Folic Acid (FA) ,etc. by CBA (Cytometric Bead Array) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 93-102% 78-104% 80-94% 90-105%
EDTA plasma(n=5) 93-105% 94-101% 86-101% 85-99%
heparin plasma(n=5) 78-99% 94-101% 82-94% 96-105%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
96-well plate 1 Plate sealer for 96 wells 4
Pre-Mixed Standard 2 Standard Diluent 1×20mL
Pre-Mixed Magnetic beads (22#:FA) 1 Analysis buffer 1×20mL
Pre-Mixed Detection Reagent A 1×120μL Assay Diluent A 1×12mL
Detection Reagent B (PE-SA) 1×120μL Assay Diluent B 1×12mL
Sheath Fluid 1×10mL Wash Buffer (30 × concentrate) 1×20mL
Instruction manual 1

Assay procedure summary

1. Add 200μL analysis buffer solution to each well of the plate for pre-wetting;
2. Preparation of standards, reagents and samples before the experiment;
3. Add 50μL standard or sample to each well, add 50μL prepared Detection Reagent A, add 10μL magnetic beads, and incubate 90 minutes protect from light at 25°C on shaker;
4. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent B, and incubate 30 minutes protect from light at 25°C on shaker;
5. Wash plate once on magnetic frame;
6. Add 100μL sheath fluid, swirl for 10 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Annals of Surgical Oncology Vitamin Status and the Development of Postoperative Cognitive Decline in Elderly Surgical Oncologic Patients. pubmed:29058145
ACS Applied Materials & Interfaces Folic Acid-Functionalized Hybrid Photonic Barcodes for Capture and Release of Circulating Tumor Cells Pubmed:29882648
Journal of Pharmaceutical Sciences Development of Orally Applicable, Combinatorial Drug–Loaded Nanoparticles for the Treatment of Fibrosarcoma Pubmed:29339136
Aging Male Role of folic acid deficiency as a possible risk factor for erectile dysfunction Pubmed: 29157083
International Urology and Nephrology Low serum folic acid can be a potential independent risk factor for erectile dysfunction: a prospective case–control study Pubmed: 30547361
Int J Biol Macromol Enhanced cellular uptake, transport and oral bioavailability of optimized folic acid-loaded chitosan nanoparticles Pubmed:35292282
Frontiers in Nutrition Folate, Vitamin B12, and Homocysteine Levels in Women With Neural Tube Defect-Affected Pregnancy in Addis Ababa, Ethiopia Pubmed:35464038
Catalog No. Related products for research use of Pan-species (General) Organism species Applications (RESEARCH USE ONLY!)
CPA610Ge11 BSA Conjugated Folic Acid (FA) Immunogen; SDS-PAGE; WB.
CPA610Ge21 OVA Conjugated Folic Acid (FA) Immunogen; SDS-PAGE; WB.
CPA610Ge41 HSA Conjugated Folic Acid (FA) Immunogen; SDS-PAGE; WB.
PAA610Ge01 Polyclonal Antibody to Folic Acid (FA) ELISA
LAA610Ge81 FITC-Linked Polyclonal Antibody to Folic Acid (FA) WB; IHC; ICC; IF.
LAA610Ge71 Biotin-Linked Polyclonal Antibody to Folic Acid (FA) WB; IHC; ICC.
CEA610Ge ELISA Kit for Folic Acid (FA) Enzyme-linked immunosorbent assay for Antigen Detection.
CCA610Ge CLIA Kit for Folic Acid (FA) Chemiluminescent immunoassay for Antigen Detection.
LMA610Ge Multiplex Assay Kit for Folic Acid (FA) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
CMA610Ge Multiplex Assay Kit for Folic Acid (FA) ,etc. by CBA (Cytometric Bead Array) FLIA Kit for Antigen Detection.
KSA610Ge11 ELISA Kit DIY Materials for Folic Acid (FA) Main materials for "Do It (ELISA Kit) Yourself".