Multiplex Assay Kit for Moesin (MSN) ,etc. by FLIA (Flow Luminescence Immunoassay)

Membrane-Organizing Extension Spike Protein

(Note: Up to 8-plex in one testing reaction)

Specificity

This assay has high sensitivity and excellent specificity for detection of Moesin (MSN) ,etc. by FLIA (Flow Luminescence Immunoassay).
No significant cross-reactivity or interference between Moesin (MSN) ,etc. by FLIA (Flow Luminescence Immunoassay) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Moesin (MSN) ,etc. by FLIA (Flow Luminescence Immunoassay) and the recovery rates were calculated by comparing the measured value to the expected amount of Moesin (MSN) ,etc. by FLIA (Flow Luminescence Immunoassay) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 89-102 97
EDTA plasma(n=5) 85-101 97
heparin plasma(n=5) 86-95 91
sodium citrate plasma(n=5) 88-96 92

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Moesin (MSN) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Moesin (MSN) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Moesin (MSN) ,etc. by FLIA (Flow Luminescence Immunoassay) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 83-90% 80-99% 83-92% 85-97%
EDTA plasma(n=5) 82-91% 79-97% 97-105% 85-93%
heparin plasma(n=5) 88-95% 90-98% 78-88% 83-102%
sodium citrate plasma(n=5) 92-99% 90-104% 98-105% 78-99%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
96-well plate 1 Plate sealer for 96 wells 4
Pre-Mixed Standard 2 Standard Diluent 1×20mL
Pre-Mixed Magnetic beads (22#:MSN) 1 Analysis buffer 1×20mL
Pre-Mixed Detection Reagent A 1×120μL Assay Diluent A 1×12mL
Detection Reagent B (PE-SA) 1×120μL Assay Diluent B 1×12mL
Sheath Fluid 1×10mL Wash Buffer (30 × concentrate) 1×20mL
Instruction manual 1

Assay procedure summary

1. Preparation of standards, reagents and samples before the experiment;
2. Add 100μL standard or sample to each well,
    add 10μL magnetic beads, and incubate 90min at 37°C on shaker;
3. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60min at 37°C on shaker;
4. Wash plate on magnetic frame for three times;
5. Add 100μL prepared Detection Reagent B, and incubate 30 min at 37°C on shaker;
6. Wash plate on magnetic frame for three times;
7. Add 100μL sheath solution, swirl for 2 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
PLoS One Presence in the Pre-Surgical Fine-Needle Aspiration of Potential Thyroid Biomarkers Previously Identified in the Post-Surgical One PubMed: PMC3765839
PLoS One.? Proteomic Profiles in Acute Respiratory Distress Syndrome Differentiates Survivors from Non-Survivors Pubmed:25290099
Bhargava, Maneesh Proteomic Studies in Acute Hypoxic Respiratory Failure Handle: 11299
Translational Research Proteomic analysis of fine-needle aspiration in differential diagnosis of thyroid nodules Pubmed:27172385
Comparative Evaluation of Membrane-Organizing Extension Spike Protein in Gingival Crevicular Fluid in Individuals with and Without Severe Chronic …
Journal of Immunology Research Moesin Is a Novel Biomarker of Endothelial Injury in Sepsis 33628853
Catalog No. Related products for research use of Homo sapiens (Human) Organism species Applications (RESEARCH USE ONLY!)
RPC642Hu01 Recombinant Moesin (MSN) Positive Control; Immunogen; SDS-PAGE; WB.
RPC642Hu02 Recombinant Moesin (MSN) Positive Control; Immunogen; SDS-PAGE; WB.
PAC642Hu02 Polyclonal Antibody to Moesin (MSN) WB; IHC; ICC; IP.
PAC642Hu01 Polyclonal Antibody to Moesin (MSN) WB; IHC; ICC; IP.
MAC642Hu22 Monoclonal Antibody to Moesin (MSN) WB; IHC; ICC; IP.
SEC642Hu ELISA Kit for Moesin (MSN) Enzyme-linked immunosorbent assay for Antigen Detection.
LMC642Hu Multiplex Assay Kit for Moesin (MSN) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.