Specificity

This assay has high sensitivity and excellent specificity for detection of Activated Leukocyte Cell Adhesion Molecule (ALCAM).
No significant cross-reactivity or interference between Activated Leukocyte Cell Adhesion Molecule (ALCAM) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Activated Leukocyte Cell Adhesion Molecule (ALCAM) and the recovery rates were calculated by comparing the measured value to the expected amount of Activated Leukocyte Cell Adhesion Molecule (ALCAM) in samples.

MatrixRecovery range (%)Average(%)
serum(n=5)92-10195
EDTA plasma(n=5)89-10498
heparin plasma(n=5)86-10495

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Activated Leukocyte Cell Adhesion Molecule (ALCAM) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Activated Leukocyte Cell Adhesion Molecule (ALCAM) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Activated Leukocyte Cell Adhesion Molecule (ALCAM) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample1:21:41:81:16
serum(n=5)91-98%82-96%78-104%83-92%
EDTA plasma(n=5)93-105%89-101%88-97%80-93%
heparin plasma(n=5)90-103%96-105%97-105%96-105%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

ReagentsQuantityReagentsQuantity
Pre-coated, ready to use 96-well strip plate1Plate sealer for 96 wells4
Standard2Standard Diluent1×20mL
Detection Reagent A1×120µLAssay Diluent A1×12mL
Detection Reagent B1×120µLAssay Diluent B1×12mL
TMB Substrate1×9mLStop Solution1×6mL
Wash Buffer (30 × concentrate)1×20mLInstruction manual1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.

GIVEAWAYS

INCREMENT SERVICES

MagazineReference
BreastActivated Leukocyte Cell Adhesion Molecule (ALCAM) in Saudi Breast Cancer Patients as Prognostic and Predictive IndicatorPubMed: 26446295
Catalog No.Related products for research use of Homo sapiens (Human) Organism speciesApplications
RPA002Hu01Recombinant Activated Leukocyte Cell Adhesion Molecule (ALCAM)Positive Control; Immunogen; SDS-PAGE; WB.
PAA002Hu01Polyclonal Antibody to Activated Leukocyte Cell Adhesion Molecule (ALCAM)WB; IHC; ICC; IP.
FAA002Hu01Anti-Activated Leukocyte Cell Adhesion Molecule (ALCAM) Polyclonal AntibodyFlow cytometry.
LAA002Hu71Biotin-Linked Polyclonal Antibody to Activated Leukocyte Cell Adhesion Molecule (ALCAM)WB; IHC; ICC.
LAA002Hu81FITC-Linked Polyclonal Antibody to Activated Leukocyte Cell Adhesion Molecule (ALCAM)WB; IHC; ICC; IF.
MAA002Hu22Monoclonal Antibody to Activated Leukocyte Cell Adhesion Molecule (ALCAM)WB; IHC; ICC; IP.
SEA002HuELISA Kit for Activated Leukocyte Cell Adhesion Molecule (ALCAM)Enzyme-linked immunosorbent assay for Antigen Detection.
SCA002HuCLIA Kit for Activated Leukocyte Cell Adhesion Molecule (ALCAM)Chemiluminescent immunoassay for Antigen Detection.