ELISA Kit for Hexosaminidase B Beta (HEXb)

Hex-B; ENC-1AS; HCC-7; Beta-N-acetylhexosaminidase subunit beta; Cervical cancer proto-oncogene 7 protein; N-acetyl-beta-glucosaminidase subunit beta

Specificity

This assay has high sensitivity and excellent specificity for detection of Hexosaminidase B Beta (HEXb).
No significant cross-reactivity or interference between Hexosaminidase B Beta (HEXb) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Hexosaminidase B Beta (HEXb) and the recovery rates were calculated by comparing the measured value to the expected amount of Hexosaminidase B Beta (HEXb) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 79-97 89
EDTA plasma(n=5) 78-90 82
heparin plasma(n=5) 82-99 93

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Hexosaminidase B Beta (HEXb) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Hexosaminidase B Beta (HEXb) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Hexosaminidase B Beta (HEXb) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 88-103% 88-102% 98-105% 89-97%
EDTA plasma(n=5) 81-93% 89-97% 90-104% 82-91%
heparin plasma(n=5) 94-102% 79-94% 80-89% 98-105%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×120µL Assay Diluent A 1×12mL
Detection Reagent B 1×120µL Assay Diluent B 1×12mL
TMB Substrate 1×9mL Stop Solution 1×6mL
Wash Buffer (30 × concentrate) 1×20mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
The American Journal of Chinese Medicine Uncovering the Molecular Mechanism of Anti-Allergic Activity of Silkworm Pupa-Grown Cordyceps militaris Fruit Body. pubmed:28367714
Life Sciences Resveratrol efficiently improves pulmonary function via stabilizing mast cells in a rat intestinal injury model pubmed:28733146
Neural Regen Res Functionality of a bicistronic construction containing HEXA and HEXB genes encoding ¦Â-hexosaminidase A for cell-mediated therapy of GM2 gangliosidoses 34100447
Nat Commun The C5a/C5a receptor 1 axis controls tissue neovascularization through CXCL4 release from platelets 34099640
Catalog No. Related products for research use of Homo sapiens (Human) Organism species Applications (RESEARCH USE ONLY!)
RPA637Hu01 Recombinant Hexosaminidase B Beta (HEXb) Positive Control; Immunogen; SDS-PAGE; WB.
RPA637Hu02 Recombinant Hexosaminidase B Beta (HEXb) Positive Control; Immunogen; SDS-PAGE; WB.
PAA637Hu01 Polyclonal Antibody to Hexosaminidase B Beta (HEXb) WB; IHC; ICC; IP.
PAA637Hu02 Polyclonal Antibody to Hexosaminidase B Beta (HEXb) WB; IHC; ICC; IP.
LAA637Hu71 Biotin-Linked Polyclonal Antibody to Hexosaminidase B Beta (HEXb) WB; IHC; ICC.
MAA637Hu22 Monoclonal Antibody to Hexosaminidase B Beta (HEXb) WB; IHC; ICC; IP.
SEA637Hu ELISA Kit for Hexosaminidase B Beta (HEXb) Enzyme-linked immunosorbent assay for Antigen Detection.
LMA637Hu Multiplex Assay Kit for Hexosaminidase B Beta (HEXb) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.