ELISA Kit for Albumin (ALB)

Serum albumin

Specificity

This assay has high sensitivity and excellent specificity for detection of Albumin (ALB).
No significant cross-reactivity or interference between Albumin (ALB) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Albumin (ALB) and the recovery rates were calculated by comparing the measured value to the expected amount of Albumin (ALB) in samples.

MatrixRecovery range (%)Average(%)
serum(n=5)83-9788
EDTA plasma(n=5)96-104101
heparin plasma(n=5)78-10384

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Albumin (ALB) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Albumin (ALB) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Albumin (ALB) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample1:21:41:81:16
serum(n=5)85-94%78-101%97-105%80-94%
EDTA plasma(n=5)78-93%82-97%85-101%96-104%
heparin plasma(n=5)87-101%98-105%83-101%89-97%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

ReagentsQuantityReagentsQuantity
Pre-coated, ready to use 96-well strip plate1Plate sealer for 96 wells4
Standard2Standard Diluent1×20mL
Detection Reagent A1×120µLAssay Diluent A1×12mL
Detection Reagent B1×120µLAssay Diluent B1×12mL
TMB Substrate1×9mLStop Solution1×6mL
Wash Buffer (30 × concentrate)1×20mLInstruction manual1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 50µL standard or sample to each well.
    And then add 50µL prepared Detection Reagent A immediately.
    Shake and mix. Incubate 1 hour at 37°C;
3. Aspirate and wash 3 times;
4. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
5. Aspirate and wash 5 times;
6. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
7. Add 50µL Stop Solution. Read at 450 nm immediately.

GIVEAWAYS

INCREMENT SERVICES

MagazineReference
PLoS OneSimvastatin Reduces Endotoxin-Induced Acute Lung Injury by Decreasing Neutrophil Recruitment and Radical FormationPubMed: 22701728
Journal of Infectious DiseasesAdjunctive corticosteroid therapy improves lung immunopathology and survival during severe secondary pneumococcal pneumonia in micePubmed: 24273183
Mol Cell BiochemThe wnt/β-catenin signaling pathway participates in rhein ameliorating kidney injury in DN micePubMed: 26346164
Catalog No.Related products for research use of Mus musculus (Mouse) Organism speciesApplications
NPB028Mu01Native Albumin (ALB)Positive Control; Immunogen; SDS-PAGE; WB.
RPB028Mu01Recombinant Albumin (ALB)Positive Control; Immunogen; SDS-PAGE; WB.
RPB028Mu02Recombinant Albumin (ALB)Positive Control; Immunogen; SDS-PAGE; WB.
PAB028Mu01Polyclonal Antibody to Albumin (ALB)WB; IHC; ICC; IP.
PAB028Mu02Polyclonal Antibody to Albumin (ALB)WB; IHC; ICC; IP.
LAB028Mu91HRP-Linked Polyclonal Antibody to Albumin (ALB)WB; IHC; ICC.
LAB028Mu71Biotin-Linked Polyclonal Antibody to Albumin (ALB)WB; IHC; ICC.
CEB028MuELISA Kit for Albumin (ALB)Enzyme-linked immunosorbent assay for Antigen Detection.
CCB028MuCLIA Kit for Albumin (ALB)Chemiluminescent immunoassay for Antigen Detection.
KSB028Mu11ELISA Kit DIY Materials for Albumin (ALB)Main materials for "Do It (ELISA Kit) Yourself".