ELISA Kit for Osteocalcin (OC)

BGLAP; OT; BGP; Bone Gla Protein; Bone Gamma-Carboxyglutamate Protein; Gamma-carboxyglutamic acid-containing protein

Specificity

This assay has high sensitivity and excellent specificity for detection of Osteocalcin (OC).
No significant cross-reactivity or interference between Osteocalcin (OC) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Osteocalcin (OC) and the recovery rates were calculated by comparing the measured value to the expected amount of Osteocalcin (OC) in samples.

MatrixRecovery range (%)Average(%)
serum(n=5)86-9894
EDTA plasma(n=5)90-9893
heparin plasma(n=5)80-9082

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Osteocalcin (OC) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Osteocalcin (OC) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Osteocalcin (OC) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample1:21:41:81:16
serum(n=5)80-93%92-99%84-98%94-105%
EDTA plasma(n=5)79-97%85-95%80-97%80-102%
heparin plasma(n=5)80-103%92-101%79-88%79-99%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

ReagentsQuantityReagentsQuantity
Pre-coated, ready to use 96-well strip plate1Plate sealer for 96 wells4
Standard2Standard Diluent1×20mL
Detection Reagent A1×120µLAssay Diluent A1×12mL
Detection Reagent B1×120µLAssay Diluent B1×12mL
TMB Substrate1×9mLStop Solution1×6mL
Wash Buffer (30 × concentrate)1×20mLInstruction manual1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.

GIVEAWAYS

INCREMENT SERVICES

MagazineReference
Acta BiomaterialiaSurface engineering of titanium with potassium hydroxide and its effects on the growth behavior of mesenchymal stem cellsPubMed: 19963080
BiomaterialsRegulation of the differentiation of mesenchymal stem cells in vitro and osteogenesis in vivo by microenvironmental modification of titanium alloy surfacesScienceDirect: S0142961212000865
International Journal of Cardiology Valvular osteoclasts in calcification and aortic valve stenosis severityPubMed: 23452891
Nephrology Dialysis TransplantationMagnesium prevents phosphate-induced calcification in human aortic vascular smooth muscle cellsPubMed: PMC3611891
Journal of Cranio-Maxillofacial SurgeryThe use of Tri-Calcium Phosphate (TCP) and stem cells for the regeneration of osteoperiosteal critical-size mandibular bony defects, an in vitro and preclinical studyPubmed: 24485270
International Journal of CardiologyValvular osteoclasts in calcification and aortic valve stenosis severity.Pubmed: 23452891
Tissue Eng Part A. Toward the Development and Characterization of an Easy Handling Sheet-Like Biohybrid Bone SubstitutePubmed:25761235
FEBS LettersLow-intensity pulsed ultrasound-induced ATP increases bone formation via the P2X7 receptor in osteoblast-like MC3T3-E1 cellsPubmed:25542352
Journal of Alloys and CompoundsEnhanced biocompatible and hemocompatible nano/micro porous surface as a biological scaffold for functionalizational and biointegrated implantsscience:S0045653516302016
Protective Effect of Soybean and Mushroom against Glucocorticoid-Induced Osteoporosis in Female Rats:
Journal of Functional FoodsComparative effects of brown and golden flaxseeds on body composition, inflammation and bone remodelling biomarkers in perimenopausal overweight womenS1756464617301561
Catalog No.Related products for research use of Homo sapiens (Human) Organism speciesApplications
RPA471Hu01Recombinant Osteocalcin (OC)Positive Control; Immunogen; SDS-PAGE; WB.
CPA471Hu21OVA Conjugated Osteocalcin (OC)Immunogen; SDS-PAGE; WB.
CPA471Hu22OVA Conjugated Osteocalcin (OC)Immunogen; SDS-PAGE; WB.
CPA471Hu23OVA Conjugated Osteocalcin (OC)Immunogen; SDS-PAGE; WB.
CPA471Hu31KLH Conjugated Osteocalcin (OC)Immunogen; SDS-PAGE; WB.
PAA471Hu01Polyclonal Antibody to Osteocalcin (OC)WB; IHC; ICC; IP.
PAA471Hu08Polyclonal Antibody to Osteocalcin (OC)WB; IHC; ICC; IP.
MAA471Hu22Monoclonal Antibody to Osteocalcin (OC)WB; IHC; ICC; IP.
SEA471HuELISA Kit for Osteocalcin (OC)Enzyme-linked immunosorbent assay for Antigen Detection.
HEA471HuHigh Sensitive ELISA Kit for Osteocalcin (OC)Enzyme-linked immunosorbent assay for Antigen Detection.
KSA471Hu01ELISA Kit DIY Materials for Osteocalcin (OC)Main materials for "Do It (ELISA Kit) Yourself".