ELISA Kit for Collagen Type III Alpha 1 (COL3a1)

COL3-A1; COL3A1; EDS4A; Collagen Type III Alpha 1; Ehlers-Danlos Syndrome Type IV Autosomal Dominant; Collagen Alpha-1(III)chain

Specificity

This assay has high sensitivity and excellent specificity for detection of Collagen Type III Alpha 1 (COL3a1).
No significant cross-reactivity or interference between Collagen Type III Alpha 1 (COL3a1) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Collagen Type III Alpha 1 (COL3a1) and the recovery rates were calculated by comparing the measured value to the expected amount of Collagen Type III Alpha 1 (COL3a1) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 78-105 101
EDTA plasma(n=5) 94-102 99
heparin plasma(n=5) 86-102 90

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Collagen Type III Alpha 1 (COL3a1) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Collagen Type III Alpha 1 (COL3a1) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Collagen Type III Alpha 1 (COL3a1) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 82-96% 95-104% 96-103% 83-98%
EDTA plasma(n=5) 81-93% 86-101% 87-95% 78-93%
heparin plasma(n=5) 94-102% 82-94% 93-102% 80-96%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×120µL Assay Diluent A 1×12mL
Detection Reagent B 1×120µL Assay Diluent B 1×12mL
TMB Substrate 1×9mL Stop Solution 1×6mL
Wash Buffer (30 × concentrate) 1×20mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
UNIVERSA MEDICINA Allogeneic human dermal fibroblasts are viable in peripheral blood mononuclear co-culture Univmed:Source
Oncotarget Epithelial but not stromal expression of collagen alpha-1 (III) is a diagnostic and prognostic indicator of colorectal carcinoma Pubmed:26741506
International Journal of Molecular Sciences Association between Plasma HMGB-1 and Silicosis: A Case-Control Study Pubmed: 30558126
Catalog No. Related products for research use of Mus musculus (Mouse) Organism species Applications (RESEARCH USE ONLY!)
RPD195Mu01 Recombinant Collagen Type III Alpha 1 (COL3a1) Positive Control; Immunogen; SDS-PAGE; WB.
RPD195Mu02 Recombinant Collagen Type III Alpha 1 (COL3a1) Positive Control; Immunogen; SDS-PAGE; WB.
PAD195Mu01 Polyclonal Antibody to Collagen Type III Alpha 1 (COL3a1) WB; IHC; ICC; IP.
PAD195Mu02 Polyclonal Antibody to Collagen Type III Alpha 1 (COL3a1) WB; IHC; ICC; IP.
LAD195Mu71 Biotin-Linked Polyclonal Antibody to Collagen Type III Alpha 1 (COL3a1) WB; IHC; ICC.
SED195Mu ELISA Kit for Collagen Type III Alpha 1 (COL3a1) Enzyme-linked immunosorbent assay for Antigen Detection.
LMD195Mu Multiplex Assay Kit for Collagen Type III Alpha 1 (COL3a1) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.